Understanding Travelling Microscope Simulator

Some lengths cannot be gripped between jaws. The bore of a capillary tube is sealed inside glass, a slit is an empty gap, and a liquid surface moves if you touch it. A travelling microscope measures such things without touching them.

The microscope does not measure anything by itself. Its job is to tell you when the cross-hair sits exactly on an edge. The length comes from the scale on the bed, which records where the carriage stood at that moment.

This is why every measurement needs two readings. One reading on its own is only a position on the bed, like a house number on a street. The size of the object is the distance between two positions, so you subtract.

The main scale is marked every 0.5 mm, and the vernier has 50 divisions that span 49 of those, a length of 24.5 mm. Each vernier division is 24.5 divided by 50, which is 0.49 mm. It falls short of a main division by 0.01 mm, and that shortfall is the least count.

A reading has two parts. The last main scale line before the vernier zero gives a value in steps of 0.5 mm. The number of the vernier line that meets a main line, times 0.01 mm, gives the rest.

Suppose the cross-hair is set on the left edge of an object. The vernier zero is just past 61.5 mm and line 27 lines up, so the first reading is 61.50 + 0.27 = 61.77 mm. At the right edge the vernier zero has just passed 64.0 mm and line 42 lines up, which gives 64.42 mm.

The width is 64.42 minus 61.77, which is 2.65 mm. Slide the object 10 mm along and both readings rise by 10 mm while the difference stays the same.

The half millimetre lines cause the most common slip. Past a half line the main scale part is 61.50 and not 61.00, so the first reading above is 61.77 mm and not 61.27 mm. This mistake does not cancel in the subtraction unless you make it on both edges.

The fine adjustment screw has a little play in its thread. If you reach one edge moving right and the other moving left, that play ends up inside the width. Start to the left of the object and make every final approach in one direction.

Sharp focus also matters. The image of the edge must lie in the plane of the cross-hairs. Move your eye from side to side, and if the edge seems to slide against the cross-hair, adjust the focus until it stays put.

Students meet this instrument when they find the bore of a capillary tube for surface tension or viscosity work. In the viscosity experiment the radius of the bore is raised to the fourth power, so a 1 percent error in the bore becomes a 4 percent error in the result. That is the reason for reading to 0.01 mm.

Many textbooks quote this scale in centimetres, where the least count is 0.001 cm and 2.65 mm is 0.265 cm. Record each reading to two decimal places before you subtract.